To order call 412-357-0373
Main Specification |
Wavelength range |
190 to 900n |
|
Wavelength accuracy |
±0.5nm | |
|
Resolution |
better than 0.3nm | |
|
Baseline stability |
0.005A/30min | |
|
Background correction |
The D2 lamp background correction capability at 1A is better than 30 times.
The S-H background correction capability at 1.8A is better than 30 times. (for WFX-120/110) | |
|
Hollow Cathode Lamps |
Lamp turret |
6-lamp turret (WFX-110/120), 4-lamp turret (WFX-130) |
|
Lamp current adjustment |
Automatic adjustment and display. Wide pulse current: 0~25mA, Narrow pulse current: 0~10mA. (WFX-110/120) | |
|
Lamp power supply mode |
Powered by two pulses (wide and narrow) simultaneously in WFX-110/120; square-wave pulse in WFX-130. | |
|
Optical System |
Monochomator |
Single beam, Czerny-Turner design grating monochromator |
|
Grating |
1800 l/mm | |
|
Focal length |
277mm | |
|
Blazed Wavelength |
250nm | |
|
Spectral Bandwidth |
0.1nm, 0.2nm, 0.4nm, 1.2nm | |
|
Detection and Data Processing System |
Detector |
Photomultiplier with high sensitivity and wide spectral range. |
|
Software |
Ms-WindowsTM operating system | |
|
Analytical method |
Working Curve Method (1st order and 2nd order fitting), Standard Addition Method. | |
|
Repeatable measurement |
Maximum 20 times of repeat measurement, automatic calculation of mean value, standard deviation and relative standard deviation. | |
|
Condition reading |
With model function | |
|
Result printing |
Measurement data and final analytical report printout. | |
|
Flame Analysis |
Burner |
10cm single slot all-titanium burner for Air-C2H2 flame(WFX-110/120/130); 5cm single slot all-titanium burner for oxygen-rich flame (WFX-110). |
|
Spray chamber |
All-plastic spray chamber. | |
|
Nebulizer |
High efficiency glass nebulizer with metal sleeve. | |
|
Position adjustment |
Adjusting mechanism for vertical, horizontal positions and the rotation angle of the burner. | |
|
Protection for gas system |
Fuel gas leak protection, abnormal flame extinction alarm, air-pressure deficiency protection and explosion-proof device. | |
|
Flame ignition |
Automatic | |
|
Characteristic Concentration and Detection Limit |
Normal Air-C2H2 flame |
Cu: Characteristic concentration ≤0.05mg/L, Detection limit ≤0.006mg/L; |
|
Oxygen-rich Air-C2H2 flame (only for WFX-110) |
Ba: Characteristic concentration ≤0.22mg/L | |
|
Hydride generation |
Hg: detection limit ≤0.5ug/L | |
|
Power Supply |
220V, 200W | |
|
Dimensions |
1020X492X537mm | |
|
Weight |
Unpacked 80kg | |